Cumulative automatic air moisture collection sampler using Peltier cooling technology


Air Moisture Collection Sampler

Air Moisture Collection Sampler

An automated air moisture collection sampler is designed and developed using thermo electric cooling technology based on the principle of Peltier effect for collecting air moisture for analysis of tritium in environment around nuclear power plants. This indigenous developed sampler collects moisture even at low humidity levels and can also be used for collection of long-time cumulative sample in comparison to the present method of random sample collection.
Air Moisture Collection Sampler

1 Air Moisture Collection Sampler

Brief Description
An automated air moisture collection sampler is designed and developed using thermo electric cooling technology based on the principle of Peltier effect for collecting air moisture for analysis of tritium in environment around nuclear power plants. This indigenous developed sampler collects moisture even at low humidity levels and can also be used for collection of long-time cumulative sample in comparison to the present method of random sample collection.

Advantages

The sampler is stand alone and doesn’t need any manual intervention for operation. It can be used in low humidity conditions and for collection of long-time cumulative air moisture sample.


Areas of Application

Air moisture collection and analysis for Tritium in different DAE units. It can also be used for analysis of any other contaminants in the air moisture.

Air Moisture Collection Sampler

1 Air Moisture Collection Sampler

Brief Description
An automated air moisture collection sampler is designed and developed using thermo electric cooling technology based on the principle of Peltier effect for collecting air moisture for analysis of tritium in environment around nuclear power plants. This indigenous developed sampler collects moisture even at low humidity levels and can also be used for collection of long-time cumulative sample in comparison to the present method of random sample collection.

Process


Being an assembly of various components, it is easy for productionisation of sampler. There is a requirement for these type of samplers in large numbers at various DAE units; Hence, there is a scope for mass production.




Advantages

The sampler is stand alone and doesn’t need any manual intervention for operation. It can be used in low humidity conditions and for collection of long-time cumulative air moisture sample.




Areas of Application

Air moisture collection and analysis for Tritium in different DAE units. It can also be used for analysis of any other contaminants in the air moisture.



System Specifications:




Sr. No.

Key Parts

Description

1

Peltier Module

Single Stage Peltier

127 P-N Junction couple

Supply: Max 12V - 6Amp

Dimension: 40 mm x 40 mm

DT max (62 ~72 degree)

Max Operating Temperature – 150~250 degree

Made of High Quality Ceramic Material

2

Heat Sink

4 copper heat Pipes

Firmly stacked Aluminium fins

Copper base with gapless touch

Integrated Cooling Fan

3

Cone

Material: Aluminium

Dia: 70 mm

Length: 138 mm

Taper Angle: 14.25°

4

Draft Tube

Material: S.S.

Large Dia: 230 mm

Small Dia: 150 mm

Length: 300 mm

Taper Angle: 7.15°

5

Temperature

Controller Unit

Multispan

12 VDC 1 Amp

6

Power Supply (SMPS)

Maxicom

240 VAC 0.5 Amp

SMPS will generate 12 V (10 Amps) DC power. The power is used to operate both Peltier unit, heat sink module and temperature controller unit.

7

UPS with batteries

ACP by Schneider electric240 VAC 5 Amp

Exide batteries 12 VDC 42 Ah X 2 nos.

External battery ups gives advantage to expand backup time in future

8

Glass tube with

sample holder and bottle

Glass tube: 12 mm 150 mm length

Sample bottle: 50 ml volume, PP.





Facilities Required

Fabrication Shop (Laser Cutting, Sheet Metal Bending, Welding), Electrical Testing and Assembly Facility, Machine Tools (Lathe, Drilling, Grinder), Manpower such as Turner, Fitter and Welder etc.



SR NO. Organization name Licensee Name Licensee Code ACQUIRED ON Valid Till Address City State Email Phone
01 Mechmatics Engineering Private Limited Devang Modi 420 23-06-2022 22-06-2027 Plot No 553 Phase 2 GIDC Vatva, Near Vatva Railway Station, Ahmedabad -382445 Ahmedabad Gujarat devang.modi@mechmatics.com 9898123425
SR NO. COUNTRY NAME PATENT NUMBER
No Data Found.
SR NO. NAME STANDARD NUMBER LOGO
No Data Found.

General License Fee : ₹ 160,000.00 (One Lakh Sixty Thousand)

General Royalty : 0%